MTL Fabrication

Tube5C-Thickox - Standard Operating Procedure


This equipment donated by Intel
CORAL Name: tube5c-thickox
Model Number: Thermco Atmospheric Furnace
Location: ICL
What it does: Atmospheric furnace
Introduction:

THERMCO “tube5C-Thickoxe” Furnace is an Atmospheric furnace designated to grow thermal Silicon Dioxide in the temperature range of 500 – 1100 °C, on Silicon wafers up to 6 inch in diameter. Other CMOS compatible thermal processes could run too. The system is equipped with an external Persys pyrogenic torch, that allows wet oxidation at temperatures bellow the pyrogenic reaction limit. This furnace has a 40” long flat temperature profile within +/- 1C. The high H2 and O2 flow capability makes the system more suitable for growing thick SiO2 films. (up to 1 um). The following gases are available:

Nitrogen 20 SLM
Oxygen 20 SLM
Oxygen 2 SLM
HYDROGEN 20 SLM (interlocked with Oxygen flow)
TCA (C2H2Cl2) 500 sccm (interlocked with Oxygen flow)

The THERMCO furnace is controlled by the TMX tube computer; while the SEMY supervisor system is
used to monitor the system and to edit and store the process recipes.
The THERMCO “tube5c-thickox” horizontal furnace is for CMOS process only.

Safety:

The system has hardware and software safety interlocks, to prevent any Hydrogen related hazard. High temperature: many furnace parts can be hot. Use caution when handling them.

Procedure:
  1. “Engage” in CORAL for ICL “tube5c-thickox” before you start. The wafers should have been RCA cleaned less than 4 hours before and your process must be approved by the PTC.

  2. Pull the tube boat out as follows:

    - On the SEMY terminal double click on “RUN” ; the “Minispec Loader” pops up on the screen.
    - From the OPTION/ TUBE menu choose “5C”
    - Click on File/Open menu , select the “ BOAT_OUT” minispec and click OK
    - On Minispec Loader window click “Compile & Send”; Compile Status window will be show up.

    Wait until the OK button is active, then click on it: the recipe was successfully sent to the TMX tube
    computer.

    Note: The tube should be in “Standby” mode to accept a new recipe; if it is in “Complete” status, push the “STOP” button on the tube TMX display to bring the system in ” Standby “ mode. Open the Tube Status from the Semy pallete; from Option/Tube select “5C”, and from Display select Detail. Check in the right top corner if the “BOAT_OUT” minispec was downloaded to the tube; the minispec listed there will be processed when you will push the START button.

    - On the tube TMX display push the “START” button; the system will go in Processing mode, and the boat will come out in 10 minutes.

  3. Pull the central quartz boat from the tube paddle Using the pick up tool, place it on the designated 5C transfer quartz plate, and place your wafers and the monitors in the quartz boat. Place the quartz boat back in the central position on the tube paddle, with the wafers front side facing the Load Zone end.

  4. Start the processing recipe, by following these steps:
    - On the SEMY terminal double click on “RUN” icon; the “Minispec Loader” pops up on the screen.
    - From the OPTION/ TUBE menu choose “5C”
    - Click on File/Open menu , select the minispec and click OK. (For example if you want to do a 1050C variable time wet oxidation select “1050C WET TUBE3 VAR” minispec)
    - On Minispec Loader window click “Compile & Send; the Compile Status window and the Variable Delay Entry window will show up. On the Variable Delay Entry window input the deposition time and push “ENTER”, input the deposition time again to confirm and wait until the OK button on the Compiler Status window is active, then click on it: the recipe was successfully sent to TMX tube computer.

    Note: The tube should be in “Standby” mode to accept a new recipe; if it is in “Complete” status, push the “STOP” button on the tube TMX display to bring the system in ” Standby “ mode. Open the Tube Status from the Semy pallete; from Option/Tube select “5C”, and from Display select Detail. Check in the right top corner if the right minispec was downloaded to the tube; the minispec listed there will be processed when you will push the START button

    - On the TMX tube display push “START” button, the system will go in “Processing” mode.

  5. When the process is finished, take the quartz boat from the tube paddle, Using the pick up tool and place it on the designated transfer quartz plate. Remove your process wafers and monitor wafers from the quartz boat. Place the quartz boat back on tube paddle and put the boat in by following these steps:

    - On the SEMY terminal double click on “RUN” icon; the “Minispec Loader” pops up on the screen.
    - From the OPTION/ TUBE menu choose “5C”
    - Click on File/Open menu , select the “BOAT_IN” minispec and click OK
    - On Minispec Loader window click “Compile & Send”: the Compile Status window shows up.

    Wait until the OK button is active, then click on it: the recipe was successfully sent to the TMX tube Computer.

    Note: The tube should be in “Standby” mode to accept a new recipe; if it is in “Complete” status, push the “STOP” button on the tube TMX display to bring the system in ” Standby “ mode. Open the Tube Status from the Semy pallete; from Option/Tube select “5C”, and from Display select Detail. Check in the right top corner if the “BOAT_IN” minispec was downloaded to the tube; the minispec listed there will be processed when you will push the START button

    - On the tube TMX display push the “START” button; the system will go in Processing mode, and the boat will start going in.

  6. "Disengage" the machine in CORAL, after you input the test data from the monitor wafers into the comments section window.
Author: Bernard Alamariu, 5/02