{"id":1441,"date":"2010-07-07T11:05:40","date_gmt":"2010-07-07T15:05:40","guid":{"rendered":"https:\/\/wpmu2.mit.local\/?p=1441"},"modified":"2010-07-07T11:05:40","modified_gmt":"2010-07-07T15:05:40","slug":"a-wireless-wearable-cardiac-monitor","status":"publish","type":"post","link":"https:\/\/wpmu2.mit.local\/a-wireless-wearable-cardiac-monitor\/","title":{"rendered":"A Wireless, Wearable Cardiac Monitor"},"content":{"rendered":"
\"Figure<\/a>

Figure 1: Block diagram of the ECG long-term Holter monitor system. The front end uses TI OPA333 and INA333 amplifiers and has a bandwidth from 0.5 Hz to 125 Hz. The 3-axis ADI accelerometer data samples at 2Hz to help correlate activity level with the ECG recordings. The battery is a 3.7-V, 600-mAh Li-Pol cell from Cameron-Sino.<\/p><\/div>\n

With the escalating costs of hospital visits, clinicians are opting to use at-home monitoring devices to diagnose patients.\u00a0 Current ECG Holter monitoring devices typically have 24-48 hour memory and battery capacity [1<\/a>]<\/sup>.\u00a0 With many patients experiencing intermittent heart problems that can occur once every week or month, the Holter monitor is not a good solution and an event recorder or loop recorder is required [2<\/a>]<\/sup>.\u00a0 However, each of these recorders can save only up to a few minutes of ECG recordings.\u00a0 This constraint leads to the loss of most of the data, which could be very important in alerting the user to the onset of future episodes.\u00a0 Therefore, we have developed a Holter monitor prototype with the goal of battery and memory capacity of two weeks.\u00a0 Figure 1 shows a block diagram of the system.<\/p>\n

\"Figure<\/a>

Figure 2: A photo of the wearable, wireless cardiac monitor. The monitor is coated in parylene C to protect it from water and sweat that may come in contact with the circuits.<\/p><\/div>\n

We based the long-term monitor prototype on a Texas Instruments MSP430 low-power microcontroller that enables high computing power with very low power consumption.\u00a0 The prototype monitor is mounted on standard 3M 2560 Red Dot electrodes and fabricated on a flexible PCB substrate.\u00a0 Mounting the PCB directly on the electrodes improves the SNR by an estimated 40 dB compared to using wired leads [3<\/a>]<\/sup>.\u00a0 The monitor is \u201cL\u201d-shaped with rounded corners and placed on the patient\u2019s chest (Figure 2).\u00a0 The \u201cL\u201d-shape enables several different ECG vectors to be recorded, depending on what the cardiologist wants to observe.\u00a0 The monitor has 1 GBit of FLASH memory, which is enough to store two weeks of data sampled at 250 Hz continuously with Huffman Encoding.\u00a0 Total power consumption of the system is approximately 2 mW.<\/p>\n


\r\nReferences
  1. D. Jabaudon, J. Sztajzel, K. Sievert, T. Landis, and R. Sztajzel, \u201cUsefulness of Ambulatory 7-day ECG Monitoring for the Detection of Atrial Fibrillation and Flutter After Acute Stroke and Transient Ischemic Attack,\u201d Stroke, J.<\/em> Amer. Heart Assoc.<\/em>, vol. 35, pp. 1647\u20131651, May 2004. [↩<\/a>]<\/li>
  2. M. A. Rockx, J. S. Hoch, G. J. Klein, R. Yee, A. C. Skanes, L. J. Gula, and A. D. Krahn, \u201cIs ambulatory monitoring for \u201cCommunity-acquired\u201d syncope economically attractive? A cost-effective analysis of a randomized trial of external loop recorders versus holter monitoring,\u201d AHJ 150(5), <\/em>pp. 1065.e1-1065.e5, Nov. 2005. [↩<\/a>]<\/li>
  3. A. Searle and L. Kirkup, “A Direct Comparison of Wet, Dry and Insulating Bioelectric Recording Electrodes,” Physiol. Meas., <\/em>vol. 21, pp. 271-283, 2000. [↩<\/a>]<\/li><\/ol><\/div>","protected":false},"excerpt":{"rendered":"

    With the escalating costs of hospital visits, clinicians are opting to use at-home monitoring devices to diagnose patients.\u00a0 Current ECG…<\/p>\n<\/div>","protected":false},"author":2,"featured_media":0,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":[],"categories":[26,27],"tags":[67,4148],"_links":{"self":[{"href":"https:\/\/wpmu2.mit.local\/wp-json\/wp\/v2\/posts\/1441"}],"collection":[{"href":"https:\/\/wpmu2.mit.local\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/wpmu2.mit.local\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/wpmu2.mit.local\/wp-json\/wp\/v2\/users\/2"}],"replies":[{"embeddable":true,"href":"https:\/\/wpmu2.mit.local\/wp-json\/wp\/v2\/comments?post=1441"}],"version-history":[{"count":2,"href":"https:\/\/wpmu2.mit.local\/wp-json\/wp\/v2\/posts\/1441\/revisions"}],"predecessor-version":[{"id":1445,"href":"https:\/\/wpmu2.mit.local\/wp-json\/wp\/v2\/posts\/1441\/revisions\/1445"}],"wp:attachment":[{"href":"https:\/\/wpmu2.mit.local\/wp-json\/wp\/v2\/media?parent=1441"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/wpmu2.mit.local\/wp-json\/wp\/v2\/categories?post=1441"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/wpmu2.mit.local\/wp-json\/wp\/v2\/tags?post=1441"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}